Description
Electrical Features:
• Power supply: AC or battery
• Voltage: 100 to 240 V
• Current: 0.3 A (115 V) to 0.16 A (240 V)
• Frequency: 50/60 Hz
• Battery Type: Rechargeable, user-replaceable polymer battery.
• Capacity: 100 ECG impressions or 3 hours of continuous display (without printing).
• Recharge Time: Approximately 3 hours after full discharge (monitor off).
• Power consumption: Approx. 40 VA
ECG Amplifier:
• Sample rate: 4,000 samples/second/channel
• Measurement range: Dynamic ± 10 mV, DC ± 300 mV
• Pacemaker detection: > ± 2 mV / pulse width> 0.1 ms
• Frequency range: 0 Hz – 150 Hz (IEC/AHA) / 0.05 Hz – 300 Hz (for children and SAECG)
• CMRR: > 100 dB
• Digital resolution: 5 μV / 18 bits
• Input Impedance: 100 MΩ
• Patient leakage current: < 5 μA
• Patient Input Circuit: Fully floating and isolated, protected against defibrillation
• Defibrillation protection: 5,000 V DC
• Dimensions: 258 x 193 x 50 mm (L/W/H)
• Weight: 1.5 kg (rechargeable battery and paper included)
Additional details
Pick-ups of equipment must be made by appointment until April 16.
If you would like to get a packaging and transport quote, please contact JMA-EXPRESS company: auction@jma-express.com