1 / 6

ARLYTE - Ionogram Apparatus

Current bid
Status

Total bid amount

See total breakdown

Max bid amount

See total breakdown
Lot specifications
Quantity
1
Margin
No
Brand
ARLYTE
Description

This device measures Sodium, Potassium, Chloride, Calcium,  and pH  using the manometric method.

Type of sample to be used:

* Whole blood, serum and/or plasma (100 μl - 150 μl)

* Pre-diluted urine (200 μl)

* Data retention: Up to  1,000 samples with results every 24 hours

Detection Range:

* whole blood, serum and plasma

* K+: 0.5 - 15.0 mmol/L

* Na+: 20 - 200 mmol/L

* Cl-: 20 - 200 mmol/L

* Ca²+: 0.1 - 5.0 mmol/L

* pH: 4.0 - 9.0 pH

Calibration of an ISE 6000 Ionogram

2 calibration points:

Cal A:

* K+: 4.0 mmol/L

* Na+: 140.0 mmol/L

* Cl-: 100.0 mmol/L

* Ca²+: 1.0 mmol/L

* pH: 7.4

Cal B:

* K+: 8.0 mmol/L

* Na+: 110.0 mmol/L

* Cl-: 70.0 mmol/L

* Ca²+: 2.0 mmol/L

* pH: 7.8

* User Interface: LCD widescreen

* Thermal Printer: 32 character

* Ambient temperature: 10° - 30° C

* Humidity: < 70%

* Serial Port: RS 232

* Voltage: 220 V +/- 10% or 110 V +/- 10%; 50 Hz +/-1 Hz

* Power Consumption: < 60W

* Dimensions: 50 cm x 27 cm x 36 cm

* Weight: 7.2 Kg

Precision:

* K+: +/- 0.2 mmol/L

* Na+: +/- 2.0 mmol/L

* Cl-: +/- 2.5 mmol/L

* Ca²+: +/- 0.1 mmol/L 

* pH: +/- 0.1

 Reproducibility of an Ionogram  (Whole Blood, Serum or Plasma)

* K+: < 1.5%

* Na+: < 1.5%

* Cl-: < 1.5 %

* Ca²+: < 3.0 %

* pH: < 1.0%

Additional details
Pick-ups of equipment must be made by appointment until April 16. If you would like to get a packaging and transport quote, please contact JMA-EXPRESS company: auction@jma-express.com
Documents

More from the same auction

View all

Download the Troostwijk Auctions app now!