Description
This device measures Sodium, Potassium, Chloride, Calcium, and pH using the manometric method.
Type of sample to be used:
* Whole blood, serum and/or plasma (100 μl - 150 μl)
* Pre-diluted urine (200 μl)
* Data retention: Up to 1,000 samples with results every 24 hours
Detection Range:
* whole blood, serum and plasma
* K+: 0.5 - 15.0 mmol/L
* Na+: 20 - 200 mmol/L
* Cl-: 20 - 200 mmol/L
* Ca²+: 0.1 - 5.0 mmol/L
* pH: 4.0 - 9.0 pH
Calibration of an ISE 6000 Ionogram
2 calibration points:
Cal A:
* K+: 4.0 mmol/L
* Na+: 140.0 mmol/L
* Cl-: 100.0 mmol/L
* Ca²+: 1.0 mmol/L
* pH: 7.4
Cal B:
* K+: 8.0 mmol/L
* Na+: 110.0 mmol/L
* Cl-: 70.0 mmol/L
* Ca²+: 2.0 mmol/L
* pH: 7.8
* User Interface: LCD widescreen
* Thermal Printer: 32 character
* Ambient temperature: 10° - 30° C
* Humidity: < 70%
* Serial Port: RS 232
* Voltage: 220 V +/- 10% or 110 V +/- 10%; 50 Hz +/-1 Hz
* Power Consumption: < 60W
* Dimensions: 50 cm x 27 cm x 36 cm
* Weight: 7.2 Kg
Precision:
* K+: +/- 0.2 mmol/L
* Na+: +/- 2.0 mmol/L
* Cl-: +/- 2.5 mmol/L
* Ca²+: +/- 0.1 mmol/L
* pH: +/- 0.1
Reproducibility of an Ionogram (Whole Blood, Serum or Plasma)
* K+: < 1.5%
* Na+: < 1.5%
* Cl-: < 1.5 %
* Ca²+: < 3.0 %
* pH: < 1.0%